Hach BioTector B3500dw User Manual page 142

Online toc
Table of Contents

Advertisement

52_HIGH CO2 IN BASE
62_SMPL PUMP STOP ON
63_SMPL PUMP STOP OFF
81_ATM PRESSURE HIGH
82_ATM PRESSURE LOW
83_SERVICE TIME
84_SAMPLER ERROR
88_O2 CONTROLLER
WARN
CO
level in the base reagent, which is
2
checked during the Zero Calibration or
Zero Check cycles, is higher than the
"BASE CO2 ALARM" level
programmed in Fault Setup menu.
Sample Pump stopped with its rotation
sensor ON or the sensor is faulty and it
is always showing ON.
Sample Pump stopped with its rotation
sensor OFF or the sensor is faulty and
it is not sensing the rotation of the
pump.
System atmospheric pressure sensor
reading is above 115kPa. Therefore,
the sensor reading is defaulted to
101.3kPa fault operation mode.
System atmospheric pressure sensor
reading is below 60kPa. Therefore, the
sensor reading is defaulted to
101.3kPa fault operation mode.
The service counter has counted down
the days between service intervals and
indicates that service is required.
Warning generated on the BioTector
venturi/vacuum sampler due to no/low
sample or low air pressure/vacuum in
the sampler.
A problem has been found during
communication with the O2-Controller
Board.
The communication between the O2-
Controller Board and the Motherboard
is lost.
Confirm the base reagent CO
filter is serviced and in good
condition, and the container is
sealed.
Check reagent quality.
Replace the base reagent.
Run the Sample Pump and check
the rotation.
Check pump sensor signal looking
at DI15 in Digital Input menu.
Replace the pump.
Run the Sample Pump and check
the rotation.
Check pump sensor signal looking
at DI15 in Digital Input menu.
Replace the pump.
Check ADC[05] in Analog Input
menu, which should be ~4 Volts.
Faulty pressure sensor.
Replace the motherboard.
Check ADC[05] in Analog Input
menu, which should be ~4 Volts.
Faulty pressure sensor.
Replace the motherboard.
Service the BioTector.
After service, acknowledge the
warning by resetting the counter
using the "RESET SERVICE
COUNTER" function in
Diagnostics, Service menu.
Check the LCD screen of the
venturi/vacuum sampler for
details.
See BioTector venturi/vacuum
sampler user manual.
Check LED 2 (L2) on the O2-
Controller Board, it should be on.
Check the 24V DC power on the
O2-Controller Board at terminals
N01 and N02.
Check the data cable (ribbon
cable) connections on the board.
Power down and power up the
BioTector to reset the system.
Replace the board.
2
Page 142

Advertisement

Table of Contents
loading

Table of Contents