9.2

BioTector Warning Event Explanation and Remedial Action

The following warning conditions do not activate the BioTector stop sequence and leave the 4-20mA signals
un-changed and do not activate the fault relay. When the system warning is rectified, acknowledge the
warning by pressing the ENTER key in the Fault Archive menu.
WARNING
21_CO2 ANL LENS DIRTY
22_FLOW WARNING – EX
23_FLOW WARNING – SO
26_PRESSURE TEST
WARN
28_NO PRESSURE TEST
CONDITION
CO
Analyzer has detected dirt on its
2
optical system.
MFC oxygen flow through the Exhaust
"EX" Valve (MV1) has dropped below
the "FLOW WARNING" level defined in
Pressure/Flow Test Program menu
8.3.4.5 Pressure/Flow Test Program
during the Pressure/Flow Test cycle.
MFC oxygen flow through the Sample
Out "SO" port (through Reactor Valve,
MV3) has dropped below the "FLOW
WARNING" level defined in
Pressure/Flow Test Program menu
8.3.4.5 Pressure/Flow Test Program
during the Pressure/Flow Test cycle.
MFC flow did not drop below the
"PRESSURE TEST WARNING" level
defined in Pressure/Flow Test Program
menu
8.3.4.5
Pressure/Flow Test
Program
during the Pressure/Flow Test
cycle.
This warning is logged when the
pressure test procedure is skipped
during the system startup sequence.
The warning cannot be acknowledged
by the user. It can only be
acknowledged automatically by the
system next time the pressure test is
passed.
CAUSE/REMEDY
Clean the CO
Analyzer.
2
Clean the lenses on the CO
Analyzer.
Various, for example,
air and oxygen supply problem.
Check oxygen pressure looking at
the Oxygen Controller Status (O2-
CTRL STATUS) menu. It should
be 400mbar (±10mbar) at 20 l/h
MFC flow.
Partially blocked ozone destructor.
Partially blocked tube after MFC.
Faulty/blocked Exhaust Valve.
Faulty MFC.
Run Flow Test (see Section
8.1.1.2 Flow Test
for details).
Various, for example,
air and oxygen supply problem.
Check oxygen pressure looking at
the Oxygen Controller Status (O2-
CTRL STATUS) menu. It should
be 400mbar (±10mbar) at 20 l/h
MFC flow.
Faulty/blocked Reactor Valve
and/or Sample Valve.
Partially blocked tube after MFC.
Faulty MFC.
Run Flow Test (see Section
8.1.1.2 Flow Test
for details).
Small gas/liquid leak in the
BioTector.
Valve leaking or not sealing.
Open Reactor Valve, inspect for
dirt/damage.
Check system fittings.
Check Mixer Reactor.
Run Pressure Test (see Section
8.1.1.1 Pressure Test
for details).
This warning occurs when the
Pressure/Flow Test is deactivated,
and when the "QUICK STARTUP"
function is used. See Section
2.2.1
Start Stop
for details.