8.3.12

Ozone Destructor Program

PWM DUTY
PURGE CYCLE MODE
MAN PURGE CYCLE
PURGE CYCLE FLOW
AUTO PURGE CYCLE
▪ PULSE WIDTH MODULATION defines the standard operation of
50%
▪ When the temperature of the BioTector exceeds 40°C for more
▪ When the temperature of the BioTector is below 15°C for more
AUTO
▪ PURGE CYCLE MODE, which is automatically controlled (AUTO
▪ When PURGE CYCLE MODE is OFF, the Ozone Destructor
▪ When PURGE CYCLE MODE is programmed as MAN (Manual),
2000, 15, 15
▪ When the PURGE CYCLE MODE above is programmed as
▪ When MANUAL PURGE CYCLE is selected, above operation
▪ This menu item is only displayed, when the PURGE CYCLE
48 l/h
▪ When the PURGE CYCLE MODE above is programmed as
▪ This menu item is only displayed, when the PURGE CYCLE
15 , 15
▪ When the PURGE CYCLE MODE above is programmed as
▪ This menu item is only displayed, when the PURGE CYCLE
the Ozone Destructor Heater. The default 50% settings mean
that the Ozone Destructor Heater is on for 50% of the time,
which is typically 5 seconds on and 5 seconds off during normal
operation.
than 2 minutes, system automatically applies a PWM DUTY
setting of 10%, which means that the Ozone Destructor Heater
is on for 10% of the time, which is typically 1 second on and 9
seconds off.
than 2 minutes, system automatically applies a PWM DUTY
setting of 100%, which means that the Ozone Destructor Heater
is on all the time. When system automatically changes the
operation of the Ozone Destructor Heater, no system warnings
are generated.
by default), specifies how the purging of the Ozone Destructor is
carried out. In AUTO (Automatic) mode, the Ozone Destructor
heater and the BioTector Cooler are controlled by the PURGE
CYCLE FLOW and AUTO PURGE CYCLE parameters defined
below.
heater and the Cooler operate normally.
the Ozone Destructor heater and the Cooler operates as
described in MAN PURGE CYCLE below.
MAN above, in MANUAL PURGE CYCLE setting, the first
parameter (2000 by default) defines the number of analysis
cycles the Ozone Destructor and the BioTector Cooler operates
as normal. The second parameter (15 by default) defines the
number of analysis cycles the Ozone Destructor heater is
switched off. The third parameter (15 by default) defines the
number of analysis cycles both the Ozone Destructor heater and
the Cooler are switched off.
routine is repeated with the relevant number of programmed
analysis cycles.
MODE operation mode is selected as MAN above.
AUTO, and if the flow measured during the Flow Test (see
Section
8.3.4.5 Pressure/Flow Test Program
the default PURGE CYCLE FLOW setting of 48 l/h, the Ozone
Destructor heater and the Cooler operates as described in
AUTO PURGE CYCLE below.
MODE operation mode is selected as AUTO above.
AUTO above, in AUTOMATIC PURGE CYCLE setting, the first
parameter (15 by default) defines the number of analysis cycles
the Ozone Destructor heater is switched off. The second
parameter (also 15 by default) defines the number of analysis
cycles both the Ozone Destructor heater and the Cooler are
switched off. The AUTO PURGE CYCLE operation routine is
carried out only once after the Flow Test routine.
MODE operation mode is selected as AUTO above.
for details) is below
Page 133