If the Stability Indicator has fewer than five bars darkened, or the
numbers on the display seem erratic, first check to make sure that there is
an adequate film of couplant beneath the transducer, and that the
transducer is seated flat against the material. If the condition persists, it
may be necessary to select a different transducer (size or frequency) for
the material being measured. See page 24.
While the transducer is in contact with the material that is being
measured, the MMX-6 will perform four measurements every second,
updating its display as it does so. When the transducer is removed from
the surface, the display will hold the last measurement made.
Occasionally, a small film of couplant will be drawn out between the
transducer and the surface as the transducer is removed. When this
happens, the MMX-6 may perform a measurement through this couplant
film, resulting in a measurement that is larger or smaller than it should be.
This phenomenon is obvious when one thickness value is observed while
the transducer is in place, and another value is observed after the
transducer is removed.
In addition, measurements through very thick paint or coatings may
result in the paint or coating being measured rather than the actual material
intended. The responsibility for proper use of the instrument, and
recognition of these types of phenomenon's, rest solely with the user of the
instrument.