Download Print this page

DAKOTA ULTRASONICS MX-1 Operation Manual page 9

Ultrasonic thickness gauge

Advertisement

The transducer is the "business end" of the MX-1. It transmits and
receives the ultrasonic sound waves which the MX-1 uses to calculate the
thickness of the material being measured. The transducer connects to the
MX-1 via the attached cable, and two coaxial connectors. When using
transducers manufactured by Dakota Ultrasonics, the orientation of the
dual coaxial connectors is not critical: either plug may be fitted to either
socket in the MX-1.
The transducer must be used correctly in order for the MX-1 to produce
accurate, reliable measurements. Below is a short description of the
transducer, followed by instructions for its use.
This is a bottom view of a typical transducer. The two semicircles of the
wearface are visible, as is the barrier separating them. One of the
semicircles is responsible for conducting ultrasonic sound into the material
being measured, and the other semicircle is responsible for conducting the
echoed sound back into the transducer. When the transducer is placed
1.888.610.7664