Aspect 100US Operator Manual PRODUCT DATA • Low attenuation carbon fiber tabletop • Tabletop: 61 cm x 117 cm (24 in x 46 in) • Imaging area: 61 cm x 117 cm (24 in x 31 in) • Height range: 66 cm to 101 cm (26 in to 38 in) •...
Aspect 100US Operator Manual INTRODUCTION Overview This manual pertains to the specified product only. It is intended for qualified medical personnel who have been trained in the use of medical equipment. It is not designed to replace or substitute for certified training in the application of this equipment. Functional capabilities and operation of the equipment described herein which can be employed in a variety of diagnostic, therapeutic, and surgical applications.
For technical assistance, call IDI at (978) 829-0009 or send fax to (978) 829-0027. Be prepared to give the complete model and serial number found on the dataplate on the table base at the time of contact.
Aspect 100US Operator Manual SYMBOLS AND DATA PLATE Intertek Testing Service The system was tested and found to be in compliance with the requirements of all relevant directives and standards in effect within the European Union at the time of manufacture. Attention! Consult accompanying documents.
Page 7
Aspect 100US Operator Manual Recycle Some of the material can be recycled rather than discarded. Not for Patient Transport The table should not be relocated with a patient on it nor should it be used to move a patient. Do Not Sit This label is found on parts that are not designed to support the sitting weight of a patient.
Aspect 100US Operator Manual SPECIFICATIONS Mode of Operation: For continuous operation with short time loading Duty Cycle: 10% (6 min/hr) Type of Equipment: Class I Type B (as defined by IEC 60601-1, UL 60601-1, EN 60601-1, CSA 601.1, IEC 60601-2-32:1994 and IEC 60601-2-46) Type B protection against electrical shock Complies with FDA standards Electrical:...
Aspect 100US Operator Manual ELECTROMAGNETIC COMPATIBILITY STATEMENT This equipment may generate and use radio frequency energy. The equipment must be installed and used according to the manufacturer’s instructions in order to avoid radio frequency interference. If this equipment generates or receives interference: •...
Aspect 100US Operator Manual SETUP Unpack No special handling is required for unpacking this equipment at the site. Conventional shipping materials are used. To remove the table from the shipping pallet, first remove the shipping brackets that secure the table to the wooded pallet using a ½” wrench. Then roll the table down a temporary ramp or lift off pallet with a forklift.
Comments and questions regarding safety should be addressed to: Customer Support Image Diagnostics, Inc. 310 Authority Drive Fitchburg, MA 01420 Or call IDI at (978) 829-0009 or send fax to (978) 829-0027. Or call toll-free at (877) 304-5434. L100-3060 REV A www.imagediagnostics.com...
Aspect 100US Operator Manual SAFETY HAZARD ALERTS Alert Circumstances for Use DANGER Indicates an imminently hazardous situation which, if not avoided, will result in death or serious injury. WARNING Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. CAUTION Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury or...
Aspect 100US Operator Manual SAFETY INSTRUCTIONS This symbol, when found marked on the equipment means: ”Attention, consult accompanying documentation.” This manual should be accessible to all personnel installing, operating, or servicing this equipment. Only a qualified technician may install or maintain this equipment. Only qualified persons may operate this equipment.
Page 14
Aspect 100US Operator Manual CAUTION Do not use table for patient transport. CAUTION During operation, listen for any unusual sounds and watch for uneven operation. CAUTION Ensure that the tabletop does not contact other equipment as it moves. CAUTION If using legholder stirrups with this table, use caution.
Aspect 100US Operator Manual ACCESSORIES • PATIENT RESTRAINT STRAPS Hook & loop fasteners Patient Restraint Strap Set Part Number C000-0571 (Two straps 3” wide) • STIRRUPS Stirrups Accessory Clamps (to mount stirrups to accessory rail ) Part Number C000-0745 (pair) Part Numbert C000-0746 (order qty 2) L100-3060 REV A www.imagediagnostics.com...
Page 16
Aspect 100US Operator Manual • ARMBOARD (OPTION) WARNING! 25lbs (11kg) Maximum Load To install clamp to tabletop: - Turn knob counterclockwise (as viewed from above) to open clamp enough to slip onto tabletop - Turn knob clockwise to tighten clamp to tabletop A –...
Aspect 100US Operator Manual ELECTRIC SYSTEM A – AC Power Cord B – Control Box (controller, batteries and battery fuse inside) C – Power On Indicator D – Safe DC Voltages E – Elevating Actuator F – Foot Control L100-3060 REV A www.imagediagnostics.com...
Aspect 100US Operator Manual CONTROLLER CONNECTIONS Pull tab to access input fuses and spares (use forceps to pull out spares) PORT DEVICE ACTUATOR FOOT CONTROL BATTERY FUNCTION When the AC power cord is plugged into the wall, the battery will be charging until the battery reaches full capacity.
Aspect 100US Operator Manual GENERAL CLEANING After each procedure, the table should be carefully cleaned The disinfectant cleaners and cleaning steps on this page apply to all the parts of the table, except the lift column. • Clean tabletop, pads, accessories, and painted surfaces with a clean cloth dampened with an approved cleaner (see list below).
Aspect 100US Operator Manual MAINTENANCE ”Qualified Technician” • All maintenance procedures should be done by an experienced technician with demonstrated knowledge and skills (electrical and mechanical) relative to this type of equipment. • This individual must have access to this manual and the proper tools. ROUTINE MAINTENANCE •...
Not made with natural rubber latex • Tabletop Pad Part Number X100-1687 • Power Cord Part Number K000-0105 To replace, pull cord straight out from control box. Power Cord (120V) Contact IDI Customer Service for other voltages TROUBLE-SHOOTING SYMPTOM POSSIBLE CAUSE ACTION...
Operator Manual DISPOSAL OF COMPONENTS IDI medical tables are made up of mostly steel, copper and aluminum parts which are easily recycled. It is recommended that some components be disassembled before disposal for recycling. The table below lists components typically found in IDI products but varies with model and options.
3. What Is Not Covered Bv This Warranty: IDI does not warrant (i) any Product or part not sold by IDI or its authorized representatives, (ii) defects caused by failure to provide a suitable installation environment...
Page 24
Agreement. 8. Warranty Terms Subject to Change: IDI reserves the right to modify the terms and conditions of its Official Warranty from time to time. The warranty terms and conditions, and IDI's obligations under such, will be determined based on the prevailing version of IDI's Official Warranty in effect at the date of purchase order.
Page 25
Purchaser (Dealer). 3. What Is Not Covered By This Warranty: IDI does not warrant (i) any Product or part not sold by IDI or its authorized Distributors or representatives, (ii) defects caused by failure to provide a suitable installation...
Page 26
Aspect 100US Operator Manual 8. Warranty Terms Subject to Change: IDI reserves the right to modify the terms and conditions of its Official Warranty from time to time. The warranty terms and conditions, and IDI's obligations under such, will be determined based on the prevailing version of IDI's Official Warranty in effect at the date of purchase order.