1. Click the
icon to open the windows device manager. Three connected AC Sources
should be displayed under 'Ports (COM & LPT)' as shown:
If they are missing or shown with an error power off and disconnect the AC Sources from
the computer. Power back on and connect AC Sources to the computer one at a time.
2. Click 'Scan' and the software will automatically connect to and assign AC sources for
multi-phase operation. There will be one master with the remaining connected AC
sources set to slave.
Setting Parameters
Set Voltage (Vrms), Phase (degrees), and Range (either 150 Vrms or 300 Vrms) for each AC
source or the whole multi-phase system. The phase value of the master must remain at 0°.
Frequency (Hz) can only be set for the whole multi-phase system.
Set parameters for each AC source individually by selecting a connected AC source listed under
the 'Setting Parameters' table. The AC source you are currently setting will be highlighted in
blue as shown:
Click 'Set' to send the parameters to the selected AC source(s). More than one AC source can be
selected and set at a time by holding the Control (Ctrl) key on your keyboard and selecting AC
sources from the list.
Output
Set parameters for the selected AC Source
50