Information Readout / External Transmission - Olympus U-CBS Instructions Manual

Optical microscope accessory coded function system
Hide thumbs Also See for U-CBS:
Table of Contents

Advertisement

4

Information readout / External transmission

This system configuration enables to acquire the information of following units.
· Information of the hole position (objective position) of the revolving nosepiece
· Information of the mirror unit position of the coded illuminator
The acquired information is transmitted to PC or the microscope digital camera through the RS-232C cable to enable
following functions.
· Implement following functions using PC by interlocking to the hole positions of the revolving nosepiece or mirror
unit positions.
(Increase of the speed of exposure time control, addition of the magnification information to acquired images,
automatic scale change according to objective, etc.)
· Enable to operate the EXPOSE button of the microscope digital camera near the coarse focusing knob of the
microscope frame by using the hand switch.
(For procedures to attach the hand switch, see page 11.)
Transmitting the acquired information to the control box (BX3-CBM) enables following functions.
· Enable to change the optical elements of the motorized condenser when objective is selected by rotating the
coded revolving nosepiece.
(The motorized condenser and control box (BX3-CBM) are required.)
Coded function system