Technical data

Function/parameter
environmental conditions
permissible temperature of the environment
run time
maximum speed n
max
minimum speed n
min
maximum RCF value at n
maximum kinetic energy
noise at maximum speed
maximum sample temperature after 30 min
permanent operation
dimensions (H x W x D)
weight without rotor
-
-
-
max
Value
indoor use
max. elevation 2000 m above sea level
max. relative humidity 80 % up to 31 °C; linearly decreasing
down to 50 % relative humidity at 40 °C.
+2 °C to +40 °C
1 min - 9 h 59 min, hold = permanent operation
-1
15 000 min
(rotor-dependent, adjustable in steps of 10)
-1
300 min
21 885
<10 kNm
< 62 dB (A)
room temperature + 15 K
315 mm x 380 mm x 475 mm
40 kg