Accessories
Mikroliter rotor 75007587 and PCR-rotor 75007569
Aerosol density applications are per-
missible only with closed container
caps!
The following steps have to be carried out:
•
Lubricate the seals before inserting them (lubricant
order no. 75003500)
•
Insert the seal (C profile) in the groove at the side
of the body of the rotor.
•
Insert the O-ring into the inner groove on the
screw-on top.
24
Loading for aerosol-tight operation
For the centrifugation of hazardous
samples you must always respect the
highest permissible sample amounts.
Aerosol-tight operation presupposes that sample ves-
sels are properly filled and the rotor lid is correctly
closed.
The vessels may generally be filled only to a point
where the sample cannot reach the vessel rim during
centrifugation. For the most commonly used vessels,
the maximum allowed volumes are listed in the table.