Error
Behavior of the
centrifuge
E-00
Motor does not start.
E-02
Rotor stops without brak-
ing to standstill.
Instrument cannot be
operated.
E-03
Rotor stops without brak-
ing to standstill.
Instrument cannot be
operated.
E-06
Rotor stops without brak-
ing to standstill.
Instrument cannot be
operated.
E-08
Rotor stops without brak-
ing to standstill.
Instrument cannot be
operated.
Possible cause(s) and measures to be taken
Motor or rotor blocked.
1. Switch instrument off and on again using the mains switch.
2. Open the lid.
3. Check whether the rotor can turn freely.
If you cannot thus relieve the malfunction, call our Service.
Internal program error in memory
Switch the instrument off and on again. If the error persists, call
our Service.
Error in speed measurement
Switch the instrument off and on again. If the error persists, call
our Service.
Communication error between keyboard and main processor.
Switch the instrument off and on again. If the error persists, call
our Service.
Overvoltage at the U/F converter
Mains voltage outside tolerance. Brake resistance defective. Call
Service if trouble persists.
Troubleshooting
51