HoMedics HL158B1-BD Instruction Manual page 29

Automatic wrist blood pressure monitor
Table of Contents

Advertisement

Troubleshooting
SYMBOLS/SYMPTOMS
Cuff Fitness
Detection Symbol
BPM cannot
communicate with
Bluetooth-enabled
device
Note: If "EP" appears on the display, just return the device to your local distributor or
importer.
CONDITIONS/CAUSES
The cuff was wrapped incorrectly
(for example too loosely or too
tightly).
Paring has not been completed.
Bluetooth function is not turn on.
The distance between BPM and
Bluetooth-enabled device is out of
transmitting range.
Use an incompatible
Bluetooth-enabled device.
Use non-Bluetooth-enabled device.
Unexpected loss of
electrical/mechanical integrity.
-28-
INDICATION/CORRECTION
Please reference "applying the
Cuff" section to wrap the cuff
correctly.
Please re-pairing the BPM and
Bluetooth -enabled device with
each other.
See the "the Bluetooth
Transmission" section to turn on
Bluetooth function.
Please make sure the acceptable
distance (≦10 meters) with each
other.
Please refer to Page15 "Bluetooth
compatibility" & Page 35 "RF
Specification"
Re-insert the batteries and try
again.