ABB FSM4000 Operating Instruction page 62

Electromagnetic flowmeter
Hide thumbs Also See for FSM4000:
Table of Contents

Advertisement

62
FSM4000 ELECTROMAGNETIC FLOWMETER | OI/FSM4000-EN REV. E
... 8 Operation
... Parameter descriptions
Submenu / parameter
Submenu
Data 70 Hz
Value range
Channel
View only
-123 µs
Zero
View only
Span >V
Span <R
Zero_2
View only
Span 2 >V
Span 2 <R
Zero_4
View only
Span 4 >V
Span 4 <R
Zero_8
View only
Span 8 >V
Span 8 <R
Zero_Pre/FIR
View only
Zero_Pre
View only
Remark
This menu does not appear with an excitation frequency of 50 Hz
or 60 Hz. Only for SE41F and SE21 transmitters.
Display of the phase deviation between signal and reference
voltage in µs.
Display of the adjustment values for zero point, span forward
and span reverse when amplification 'High' is selected in the
'Service / Primary Adjust / Amplification' menu.
Display of the adjustment values for zero point, span forward
and span reverse when amplification 'Middle High' is selected in
the 'Service / Primary Adjust / Amplification' menu.
Display of the adjustment values for zero point, span forward
and span reverse when amplification 'Middle Low' is selected in
the 'Service / Primary Adjust / Amplification' menu.
Display of the adjustment values for zero point, span forward
and span reverse when amplification 'Low' is selected in the
'Service / Primary Adjust / Amplification' menu.
Display of the zero point offset with activated filter in 'Service /
Noise Handling / Prefilter /' and 'Service / Noise Handling /
Prefilter / FIR EA'
Display of the zero point offset with activated filter in 'Service /
Noise Handling / Prefilter /'