Summary of Contents for Human COMBILYZER Plus

  • Page 1 User Manual !"#$%&'()*+,-./ Cat.-No.: 17601P...
  • Page 2: Table Of Contents

    CONTENTS INTRODUCTION Methodology of Urine Testing Test Strips DESCRIPTION OF COMBILYZER PLUS INSTALLATION OPERATION GUIDE Control Menu System Setup Procedure (Menu/Setup) 4.2.1 Setting the Clock 4.2.2 Setting Serial Output Port 4.2.3 Setting the Language Strip Setup Options (Menu/Strip) 4.3.1 Setting Measurement Unit 4.3.2 Setting the Sensitivity Attribute of the Image Processing...
  • Page 3 This page left blank 2/22 User Manual COMBILYZER Plus...
  • Page 4: Introduction

    However, misreading or misinterpreting of results may happen, caused by individual handling habits of different users or different light conditions. Test Strips The COMBILYZER Plus works with COMBINA 11S urine test strips: Tested parameters are as follows: Bilirubin Urobilinogen...
  • Page 5: Description Of Combilyzer Plus

    The testing corresponds with the reaction time of the chemical reagents - about 60 seconds. The COMBILYZER Plus works using the reflectance photometry method: inside the reader, the strip is illuminated by white light, and the light reflected from the strip is detected by a colour CCD sensor.
  • Page 6 Printer Station. The instrument is designed specifically for in vitro diagnostic professional use and to meet the requirements of medical professionals. The measurement capacity is up to 80 test strip evaluations per hour. User Manual COMBILYZER Plus 5/22...
  • Page 7: Installation

    3 INSTALLATION The COMBILYZER Plus is shipped in a cardboard box. Prior to unpacking, clear the area where the instrument is to be operated: a 30 x 50 cm size space will be needed. Please note the shipping marks on the box while handling.
  • Page 8 Now, the COMBILYZER Plus is ready to operate using the Main Menu options (see chapter 9). Figure 4. Loading the printer paper Note: After installation check whether the instrument is operating accurately.
  • Page 9: Operation Guide

    4 OPERATION GUIDE Control Menu System The operation and functions of the COMBILYZER Plus can be viewed in the menu system chart (see Appendix). Each selected menu allows the user to control one of the operational or setup functions. For a detailed explanation of the submenus, please see the detailed descriptions in following chapters.
  • Page 10: Setting Serial Output Port

    German or Hungarian. The language can be selected with the UP and DOWN buttons or by directly marking the field next to the desired language. Store the changes and return to main menu by pressing OK. User Manual COMBILYZER Plus 9/22...
  • Page 11: Strip Setup Options (Menu/Strip)

    OK button. Note: When the reader has been adjusted to settings different from the standard HUMAN settings this will be indicated by asterisks in the footer of the printout. 10/22 User Manual COMBILYZER Plus...
  • Page 12: Setting Pad Sequence In The Printout

    Store the changes and return to main menu by pressing the OK button. User Manual COMBILYZER Plus 11/22...
  • Page 13: Service Menu Options (Menu/Service)

    Service Menu Options (Menu/Service) 4.4.1 Code Card Entering encoded test strip related information allows the COMBILYZER Plus to control the measurement process precisely. This coded information describes the maximum number of measurements being allowed with the given code . When opening a new vial of strips, select the Menu/Service/Calibration submenu from the menu system.
  • Page 14: Selecting Measuring Mode

    55 seconds before inserting them into the COMBILYZER Plus and pressing Start. If the reaction ( incubation ) time is too short or too high false results my be obtained. Refer also to the package insert of the COMBINA 11S test strips.
  • Page 15 9. When you are finished working with the COMBILYZER Plus press the Quit button in main menu. After asking for confirmation of your request, the instrument will shut down automatically.
  • Page 16: Memory/Data Management

    PRINT. For example on figure 10. the low limit is 39 and 41 is the high limit of the range to be printed. Pressing the PRINT button will therefore print three records, 39-40- User Manual COMBILYZER Plus 15/22...
  • Page 17: Sending Data To Host Computer

    CLEAR button. The address pointer returns to the next lower address field after execution of the memory clear function. By pressing CLR.ALL button all, existing records in the memory will be cleared. 16/22 User Manual COMBILYZER Plus...
  • Page 18: Printer Station

    In this case test results can only be viewed on the LCD. - The PC unidirectional protocol does not apply control sum checking. - When sending/uploading result data to a computer, the COMBILYZER Plus needs to be connected to the host computer (PC) via the interface cable (included) and the serial output line needs to be enabled.
  • Page 19: Maintenance

    When batteries are low a „Battery low” message appears on the LCD. In case the screen hasn't been touched for a while, the COMBILYZER Plus will shut down to save energy; after 2,5 minutes if operated on batteries and after 5 minutes if connected to mains.
  • Page 20: Error Messages And Troubleshooting

    11. Need calibration code! enter a valid calibration code Note: If a problem cannot be solved or a malfunction is experienced that is not listed above, please, contact your HUMAN distributor immediately. User Manual COMBILYZER Plus 19/22...
  • Page 21: Technical Features

    8 PACKING LIST COMBILYZER Plus instrument 1 pc Power adapter with power cord 1 pc Rechargeable battery 3 pcs Strip tray 1 pc Serial communication cable 1 pc Printer paper 1 roll User manual 1 pc 20/22 User Manual COMBILYZER Plus...
  • Page 22: Symbols

    9 SYMBOLS according to EN 980 In vitro diagnostic medical device Warning Information inside Serial number Manufacturer User Manual COMBILYZER Plus 21/22...
  • Page 23: Appendix

    10 APPENDIX The Menu Structure 22/22 User Manual COMBILYZER Plus...
  • Page 24 Human Gesellschaft für Biochemica und Diagnostica mbH Max-Planck-Ring 21 ! D-65205 Wiesbaden Germany Telefon: +49 6122 9988 0 Telefax: +49 6122 9988 100 eMail: human@human.de Internet: http://www.human.de 02/2005-05...