Summary of Contents for Human HUMAX 4K

  • Page 1 | User Manual Cat.-No. 15210/100...
  • Page 2 DATE / Rev. REVISION DESCRIPTION 01/2007-01 First edition...
  • Page 4: Intended Use Of The Instrument

    The HUMAN representative shall be relieved of any liability under this warranty if the product is not used in accordance with the manufacturer's instructions, altered in any way not specified by HUMAN, not regularly maintained, used with equipment not approved by HUMAN or used for purposes for which it was not designed.
  • Page 5: Instrument Disinfection

    Every effort has been made to avoid errors in text and diagrams, however, HUMAN GmbH assumes no responsibility for any errors which may appear in this publication. It is the policy of HUMAN GmbH to improve products as new techniques and components become available. HUMAN GmbH therefore has to reserve the right...
  • Page 6 NOTICE Analytical instruments for in vitro diagnostic application involve the handling of human samples and controls which should be considered at least potentially infectious. Therefore every part and accessory of the respective instrument which may have come into contact with such samples must equally be considered as potentially infectious.
  • Page 8 Description Installation 2.1 Contents of Package 2.2 Mains Supply 2.3 Environmental Conditions 2.4 Positioning 2.5 Inspection 2.6 Loading 2.7 Rotor Installation 2.8 Manual Lid Opening Specifications Accessories 4.1 Safety Interlock System Instructions for Use 5.1 Schematic Drawing 5.2 Controls and Indicators 5.3 Preparing The Run 5.4 Programming 5.5 Program Recall...
  • Page 9 The HuMax 4K ensures safe and easyoperation by means of a locking system, which does not allow the centrifuge to operate or the rotor to spin if the lid is not closed.
  • Page 10: Installation

    1 user manual 1 power cable 1 manual lid opening tool The HuMax 4K requires 230 V, 50 Hz. The centrifuge should be connected only to a grounded outlet. The instrument is designed to operate safely under the following conditions: Indoor use only Ambient temperature: 5 °...
  • Page 11 17157 7 ml vacuum / non-vacuum tubes The HuMax 4K Centrifuge is equipped with an interlock system that prevents the centrifuge lid from opening when the rotor is spinning. The centrifuge will not operate until the lid is closed completely. The lid remains closed until the rotor stops spinning.
  • Page 12 5 Instructions for Use Figure – 1 5/10 User Manual...
  • Page 13 Figure – 2 Speed Display This display shows the speed values during programming. It also indicates whether the instrument is accelerating (“Acc”), braking (“br”), power failure (“E oFF ”), open lid (“oPEn”) and (“cLoSE Lid”) when start is pressed while the lid is open. The error codes that occur in the case of a malfunction are also shown on this display.
  • Page 14 It occurs in case of a power failure during the run. This message will disappear after several minutes or after the lid is opened and closed again. In the case of any persistent(?)error, please contact your local HUMAN representative. 6 Hazards, Precautions And Limitations Of Use The following precautions must be observed: Never try to bypass the safety lid lock while the rotor is spinning.
  • Page 15 High voltage is present behind the panels of the centrifuge. Caution: Do not open the instrument! There are no user-servicable parts inside. In the case of a malfunction, please contact your local HUMAN representative. Use only rotors and accessories designed for use in the centrifuges.
  • Page 16 7 Cleaning And Preventive Maintenance CAUTION: All cleaning should be performed with the centrifuge disconnected from the power outlet. No daily cleaning is required, except in the case of accidental glass breakage or a large amount of spillage in the bowl.
  • Page 17 8 Electrical Circuit Diagram 10/10...