After completion of the installation process, the SONO-VIEW will re-start, verify the connected probes, and will
immediately commence to call up the measurement values.
4.2
Measurement Value Display
The SONO-VIEW immediately commences to call up and display the measurement values after start-up. This is
performed in a 500 ms cycle. Depending on how many probes are connected, from one to four, the following
displays will be presented.
If between five and eight probes are connected, a second display with four values apeers.
The moisture value in percent and the respective probe number is always presented. If two or three probes are
connected, the temperature measured by the probe is additionally also presented. In the event that only one
probe is connected, the calibrated radar run-time is also presented.
If more than one probe is connected to the SONO-VIEW, there is an option to change the display. For this purpose,
press the buttons
/
.
With each press, all connected probes are displayed one after the other as individual probes. If the last probe is
displayed as a single probe, the next time the button is pressed, it will be displayed again the display of 4 probe
values.
The probe number always appears in the upper right area of the display. If you remain on a display for a longer
time, this display is set as the "default". After a restart, the SONO-VIEW starts on the set measurement display.