(Please read this manual carefully to achieve the best and consistent column performance for a long time)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the package and outside of the column for any damage.
(2) Please check if product name and serial number written on the column package, column name tag, and
enclosed Certificate of Analysis (CoA) are matching and correct.

1. Introduction

Thank you for purchasing the Shodex product. Shodex GPC LF-804 is a size exclusion chromatography
column used with organic solvents. It is packed with unique multi-pore gels with a wide pore-size distribution.
Therefore it can be obtained highly linear calibration curve without inflection points. The column is suitable for
analysis over a broad range of molecular weights.

2. Column Components

3. Column Specifications

Product
Product
Code
Name
F6021041 GPC LF-804
F6709621 GPC LF-G
Base Material
: Spherical porous particles of styrene divinylbenzene copolymer
Column Housing : SUS-316
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : Tetrahydrofuran (THF)
Operation Manual
Shodex
GPC LF-804
Column Size
Particle
(mm)
Size (μm)
I.D.
Length
8.0
300
6
4.6
10
6
SHOWA DENKO K.K. (https://www.shodex.com/)
Theoretical
Target Molecular
Plate Number
Weight Range
(Per Column)
≥ 17,000
300 - 2,000,000
(Guard Column)
*
Reference value only / Measured with polystyrene
ver.211001E
Exclusion
*
Limit
*
2,000,000
-
-
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GPC LF-804 and is the answer not in the manual?

Questions and answers

Summary of Contents for Shodex GPC LF-804

  • Page 1 Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex GPC LF-804 is a size exclusion chromatography column used with organic solvents. It is packed with unique multi-pore gels with a wide pore-size distribution.
  • Page 2 · Shear degradation is more likely to occur in larger molecular weight compounds. When shear degradation occurs, the measured molecular weight may be lower than the actual value or it may influences the repeatability of the analysis. If shear degradation is suspected, use a lower flow rate. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 3 · If the eluent left in the system is not compatible with the column to be used, it may damage the column. · A drastic change in the eluent compositions may remove substances adsorbed on the system and they may enter and deteriorate the column. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 4 When replacing THF to DMF with LiBr, first follow step (3) above, and then replace 100 % DMF to DMF with LiBr. Attention! · Frequent solvent replacement deteriorates the column, and thus not recommended. SHOWA DENKO K.K. (https://www.shodex.com/)
  • Page 5 (2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
  • Page 6 11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.

This manual is also suitable for:

Gpc lf-gF6021041G6709621

Table of Contents