Page 1
Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex HILICpak VN-50 series is a polymer-based HILIC column modified with diol group. The column is suitable for the analysis of hydrophilic compounds such as oligosaccharides and oligonucleic acids.
Page 2
Pressure (MPa) pH Range Range (°C) Recommended Maximum (Per Column) HILICpak VN-50 2D 0.1 - 0.3 2 - 13 4 - 60 HILICpak VN-50G 2A Usable solvents are listed below. (1) A mixture of water and acetonitrile or water and methanol of any ratio can be used. Up to 100 % water, acetonitrile, or methanol can be used.
Page 3
This may result in creating an empty space in the column and deteriorates the column. Note · It is recommended to set the pump limiter to avoid exceeding the maximum pressure. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 4
(a cool and dark space is recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. Attention! · Never allow inside of the column to dry. It can damage the column. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 5
(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
Page 6
11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.