(Please read this operation manual carefully to achieve the best and consistent column performance for a long time.)
Important Handling Instructions
Caution!
· Please consult the Safety Data Sheet (SDS) of reagents and solvents used with the column
and understand their proper handling methods to prevent potential health hazards or death
from occurring.
· Please wear appropriate personal protective equipment such as lab goggles and gloves
when handling organic solvents and acid and alkaline reagents. Avoid any direct physical
contact to prevent chemical injuries.
Before Using the Column
(1) Please visually inspect the column package and the column surface for any damage.
(2) Please check the product name and serial number (Serial no. or S/N) written on the column package and
adhesive label on the column body.
(3) Please download the Certificate of Analysis (CoA) for the purchased product. The CoA can be downloaded
from Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number.
1. Introduction
Thank you for purchasing the Shodex product. Shodex KW400-4F series is silica-based gel filtration
chromatography columns. The column series is suitable for size exclusion separations of proteins. The columns
with different exclusion limits are available. Please select columns with target molecular weight ranges (see
below) that meet the molecular weight range of your samples.
2. Column Components
Please refer to the Shodex website: https://www.shodex.com/en/da/07.html
3. Column Specifications
Product
Product Name
Code
KW402.5-4F
F6989201
F6989202
KW403-4F
F6989203
KW404-4F
F6989204
KW405-4F
F6700132
KW400G-4A
Base Material
: Spherical porous particles of silica modified with hydrophilic group
Column Housing : SUS-316
Screw Type
: Internally-threaded type No.10-32 UNF
Shipping Solvent : Water
Operation Manual
Shodex KW400-4F series
Column Size
Particle
(mm)
Size
(μm)
I.D.
Length
4.6
300
3
4.6
300
3
4.6
300
5
4.6
300
5
4.6
10
5
Resonac Corporation (https://www.shodex.com/)
Theoretical
Target Molecular
Plate Number
Weight Range
(Per Column)
≥ 35,000
5,000 - 70,000
≥ 35,000
10,000 - 500,000
≥ 25,000
30,000 - (4,000,000)
≥ 25,000
200,000 - (20,000,000)
(Guard Column)
*1
Reference value only / Measured with proteins
ver.20221202E
Exclusion
*1
Limit
*1
150,000
600,000
(4,000,000)
*2
*2
*2
(20,000,000)
*2
-
-
*2
Estimated value
1

Advertisement

loading
Need help?

Need help?

Do you have a question about the KW400-4F Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Shodex KW400-4F Series

  • Page 1 Shodex website (https://www.shodex.com/download/). You will be asked to enter the serial number. 1. Introduction Thank you for purchasing the Shodex product. Shodex KW400-4F series is silica-based gel filtration chromatography columns. The column series is suitable for size exclusion separations of proteins. The columns with different exclusion limits are available.
  • Page 2: Eluent Preparation

    (1) Degas the eluent fully to prevent the formation of air bubbles. (2) Presence of small debris or insoluble substances may result in deterioration of columns and/or they may appear as noise on chromatograms. Filter the eluent with a 0.45-μm disposable filter to prevent the problems from occurring. Resonac Corporation (https://www.shodex.com/)
  • Page 3: Sample Preparation

    · Use of a guard column is recommended to protect the analytical column. 7. Column Usage Procedure KW400-4F series is semi-micro columns. To achieve its best performance, please use it with a semi-micro type HPLC system. 7.1 HPLC System Preparation Wash the entire HPLC system prior to column installation, including all flow-lines and sample loop by switching the valves, and then replace the washing solution with the eluent to be used.
  • Page 4 During the column cleaning, disconnect the detector and collect the washing solution directly from the column outlet into a waste container (i.e., do not let the solution go through the detector). Resonac Corporation (https://www.shodex.com/)
  • Page 5 · Never allow inside the column to dry. It can damage the column. 9. Column Inspection Please refer to the inspection method described in the CoA. At Shodex, “half width method” is adopted for the calculation of plate count and “asymmetry factor (Fas)” is adopted for the calculation of peak symmetry. Please refer to the Shodex website for the detail: https://www.shodex.com/en/da/07.html...