5
Remove Rear Enclosure by gently prying with a small screwdriver or equivalent at the two locations
indicated in
This will free it from being retained by the Display hinges.
6
Once the two locations are free. Grab the Rear Enclosure with both hands and pull it free. It will still be
held by one tab on each end, but they will release with a small amount of force.
7
Feed the wireless USB cable through the Rear Enclosure to complete the removal.
8
Use a T6 bit to Remove the 2 screws securing the Display Cable EMI Shield per
9
Disconnect the Clinical Display Cable from the Main PCBA per
10 Move the Clinical Display to the lowered or closed position.
11 Use a T20 bit to remove the 2 screws in each hinge per
12 Carefully lift the Clinical Display and feed the Clinical Display Cable through the Bottom Enclosure to
complete the removal.
Caution:The following steps remove the Hinge Tubes. The display cable is housed inside so extra care is
required to protect the cable.
13 The Hinge Tubes can be removed for transfer to a new display if needed per
14 Remove the center Hinge Tube first.
15 Remove the right and left Hinge Tubes.
Figure 6.8 Rear Enclosure Removal
52
Chapter 6: Replacement Procedures
Figure 6.8 on page
52. Apply pressure to move the Rear Enclosure in a downward direction.
Figure 6.10 on page
Figure 6.10 on page
Apply upward force on tool in both
locations to dislodge Rear Enclosure
from system
Pull Rear Enclosure from system. The remaining
locking tabs on the ends are also on the bottom
edge of the Enclosure
Figure 6.9 on page
53.
53.
53.
Figure 6.11 on page 54