Guidance and Manufacturer's Declaration - Electromagnetic Immunity per IEC 60601-1-1-2:2014
The Sonosite PX ultrasound system is intended for use in the electromagnetic environment
specified below.
Immunity
Test
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
Power
Frequency
Magnetic Field
IEC 61000-4-8
Conducted RF
IEC 61000-4-6
22
Chapter 3: Safety
IEC 60601 Test
Compliance
Level
Level
0% U
) for 0.5 cycle
0% U
T
cycle
0% U
) for 1cycle
T
0% U
70% U
(30% dip in
T
70% U
U
) for 500msec
T
in U
500msec
<5% U
(>95% dip in
T
U
) for 5s
T
<5% U
dip in U
3 A/m
3 A/m
3 Vrms
3 Vrms
150 kHz to 80 MHz
6Vrms in ISM bands
6 Vrms in ISM
bands
Electromagnetic
Environment
) for 0.5
Mains power quality should
T
be that of a typical
commercial or hospital
environment. If the user of
) for 1cycle
T
the Fujifilm Sonosite
ultrasound system requires
(30% dip
continued operation during
T
power mains interruptions, it
) for
T
is recommended that the
Sonosite PX ultrasound
system be powered from an
(>95%
T
uninterruptible power supply
) for 5s
or a battery.
T
If image distortion occurs, it
may be necessary to
position the Fujfilm Sonosite
ultrasound system further
from sources of power
frequency magnetic fields or
to install magnetic shielding.
The power frequency
magnetic field should be
measured in the Intended
installation location to assure
that it is sufficiently low.
Portable and mobile RF
communications equipment
should be used no closer to
any part of the Sonosite PX
ultrasound system including
cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended Separation
Distance
d = 1.2
P