Data Output Format $Pvhy2 (Post November 2020) - Valeport Hyperion Chlorophyll a Operating Manual

Table of Contents

Advertisement

6.2 Data Output Format $PVHY2 (Post November 2020)

The Hyperion outputs a single NMEA style data string – post November 2020 (Serial No: 75100)
this will be a $PVHY2 format string.
Example - Turbidity: $PVHY2,01,8,0.48705,0.012,NTU,141.65588,0.023,NTU,M1,*6D
Example – Fluorometer: $PVHY2,01,1,0.816,0.012,ug/l,,,,C02,*24
Please note that fields with no data e.g. 2
delimited, as above.
Where:
Field
Number Description
1
NMEA Header
2
Instrument ID
3
Parameter ID
st
4
1
Parameter Mean
st
5
1
Parameter SD
st
6
1
Parameter Units
nd
4
2
Parameter Mean
nd
5
2
Parameter SD
nd
6
2
Parameter Units
7
Operating Mode
8
Check Sum
If the sampling that has been setup such that only one measurement is used to produce a reading
then no Standard Deviation is generated and the output becomes:
Turbidity - $PVHY2,01,8,0.48705,,NTU,1.65588,,NTU,M1,*6D
Fluorometer - $PVHY2,01,1,0.816,,ug/l,,,,C02,*24
Otherwise:
Turbidity - $PVHY2,01,8,0.48705,0.012,NTU,141.65588,0.023,NTU,M1,*6D
Fluorometer - $PVHY2,01,1,0.816,0.012,ug/l,,,,C02,*24
© 2022 – Valeport Ltd
nd
parameter data in a fluorometer will be empty but
Type
Description
String
Valeport HYperion2
Integer
Identifies the instrument
Integer
1 = Chlorophyll a
2 = Fluorescein (Uranine)
3 = Rhodamine\Sulforhodamine B
4 = Crude Oil
7 = Phycocyanin (freshwater Blue Green Algae)
8 = Turbidity
Float
Float
String
ug/l for Chlorophyll a
NTU for turbidity nephelometer
Float
Float
String
NTU for turbidity backscatter
String
C02 (default)
Char
An exclusive OR sum between all characters
between the '$' and the '*'of the string
Section 6 | Communications
Page | 20

Advertisement

Table of Contents
loading