CLEANING

Focusing Sleeve

For best results, use the disposable sheath model #52640.
1. Remove the focusing sleeve from the Fiberoptic Light Pipe or
Fiberoptic Light Cable by rotating it counterclockwise to
unthread.
2. Use a cotton swab soaked in warm water and mild detergent
to clean the distal end of the Fiberoptic Light Pipe or Light
Cable, the proximal side of the Focusing Sleeve (small end),
and distal window on the large end on the Focusing Sleeve.
3. Dry using a new cotton swab.
4. Reassemble the Focusing Sleeve to the Light Pipe or Light
Cable by rotating clockwise.
WARNING: Do not clean any part of the focusing sleeve with
alcohol. Clean only with a mild detergent and water.
NOTE: The focusing sleeve should not be sterilized or immersed.

Fiberoptic Light Pipe, Light Cable, Neonate Light Pipe

For best results, use the disposable sheath model #52640.
External surfaces may be cleaned with a solution of warm water
and mild detergent.
NOTE: The Fiberoptic Light Pipe, Light Cable, and Neonate Light
Pipe should not be sterilized or immersed.