SET -UP AND OPERATION

Set Up

1. Secure the Light Box to one of the mounting options per the
appropriate instructions.
2. Insert the Fiberoptic Light Pipe, Fiberoptic Light Cable, or
Neonate Light Pipe into the light pipe receptacle on the top of
the Light Box.
3. Plug the power cord into the base of the Light Box and the
remaining end into a grounded AC outlet.
4. Toggle the light switch to the "ON" position to operate.
NOTE: Remember that the Fiberoptic Light Pipe, Fiberoptic Light
Cable, and Neonate Light Pipe contain several thousand individual
optical fibers, therefore appropriate handling is required to ensure
long fiber life. Do not bend the flexible portions of the light pipe at
acute angles as this will break individual fibers and reduce light
transmission. For example, do not bend the semi-flexible middle
portion of the Fiberoptic Light Pipe more than 180 degrees and do
not bend the proximal and distal sections more than 90 degrees.
Be careful not to pull on the Fiberoptic Light Cable as this may
cause fibers to break.