M&C PMA Series Instruction Manual page 34

Hide thumbs Also See for PMA Series:
Table of Contents

Advertisement

For dismounting the measuring cell, the following procedure is recommended:
Warning
Warning
 Loosen the inside hex screws of the housing window;
 Unscrew the window lid of the analyzer;
 Loosen the fastening screw (M4 in the middle left) of the inside front plate;
 Pull out the front plate and swing it out to the right side;
 Remove carefully the black isolating cap from the transmitter unit;
Warning
 Loosen the threaded tube couplings
Warning
 Pull off the green 2-, 3- and 4-pole plug-in connections (on terminal X1, X2 and X3) from the power supply board;
 Loosen the earth connection (green-yellow) of the transmitter unit
 Loosen the four hex screws
 Loosen union nuts
5
at sample in-
Now, the complete transmitter unit can be taken out of the housing. All further procedures should be executed on a
clean work bench outside the hazardous area. Put the transmitter unit in a position as shown in Figure 8 (18-pin plug
must show to the right side).
34
During all work on the analyzer it must be ensured that both the working environ-
ment and the analyzer itself are free of explosive or flammable gases!
Dangerous voltage. Before performing any maintenance work, disconnect the an-
alyzer and all external circuits connected to the analyzer from the power supply.
Hot transmitter surface up to 55 °C [131 °F].
Touching may lead to burnings.
Wear safety gloves!
5
of the sample gas inlet and the sample gas outlet (see Figure 7);
Strong magnetic field. Before dismounting the transmitter unit, remove all sensi-
ble parts (eg. wrist watch etc.)!
3
(see Figure 7) on top of the transmitter mounting plate;
8
and outlet
6
(see Figure 7);
PMA 50 EX | 1.01.04
(2
see Figure 7);
www.mc-techgroup.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pma 50 ex1803436

Table of Contents