18.1

ZERO CALIBRATION

To do a zero-point calibration of the analyzer, use an O2-free gas, e.g. nitrogen (N
below:
 Connect the flexible PVC or Viton tube with the bottle's pressure reducer of the N
Warning
First of all, open the bottle valve, then open the outlet valve of the pressure regulator;
 The pressure regulator and the tubing have to be purged for about 5 seconds;
 Check adjusted control pressure and reduce to ≤ 1.1 bar abs. if necessary, then shut-off the outlet valve of
the pressure regulator again;
 Connect the free tube end of the zero gas bottle connection to the external flow meter or, if available, to
the external calibration valve;
 Open the outlet valve of the pressure regulator slowly, in order to avoid pressure peaks;
 Adjust the zero gas volume flow with the needle valve of the flow meter to max. 60 l/h. The volume flow
of the calibration gas should always be adapted to the sample gas volume flow;
Always calibrate at the same gas flow as used for the measurement.
Note
 Wait approx. 30 seconds until display is stable;
 Set the range switch
 If necessary, adjust the zero point to 0 % via the zero point potentiometer
 Check the analogue output signals at 0% O
Ouput signal
0-10 V DC
0-20 mA
4-20 mA
If a gas mixture is analyzed, the individual components shall be checked for possible
cross-sensitivity and taken into account accordingly during zero calibration (see chapter
20).
Note
Shut the output valve of the pressure regulator and bottle valve and disconnect the tube connection to the
analyzer.
24
The pressure reducer should have a maximum outlet measuring range of 0 to
1.5 bar abs. and should always be adjusted to a low outlet pressure of max.
1.1 bar abs. This protects the measuring cell against destruction due to high
pressure!
8
(see Figure 1) to 0-1 % O
;
2
Signal to be measured
0 V
0 mA
4 mA
PMA 50 EX | 1.01.04
(with a screw driver);
2
) 5.0. Follow the steps
2
-zero gas bottle;
2
7
(Figure 1) (with a screw driver);