Electrical connections meas. cell
Gas outlet measuring cell
Gas inlet measuring cell
Fastening screw for measuring cell
Figure 8
Transmitter unit
Unsolder the brown and yellow cable from the terminals  (see Figure 8) on the back side of the measuring cell;
do not overheat the terminals; mark the cables accordingly;
Disconnect the tubing for the sample gas outlet  and sample gas inlet ;
Loosen the fastening screw of the measuring cell  with a screw driver and pick carefully the cell;
Exchange only with measuring cells of the same type;
Turn the transmitter as shown in Figure 9 and loosen the fastening screw of the photocell .
Install the measuring cell in reverse order; make sure the dumbbell is in the correct position!
In case there are minimally different positions of the dumpbells inside the measuring cells when mounting a
new cell, it is absolutely necessary to adjust the zero point mechanically.
www.mc-techgroup.com
Heater element
Transmitter board
Temperature fuse (cut-off at 72 °C [161.6 °F])
Temperature sensor
PMA 50 EX | 1.01.04