TECHNICAL DATA

05A1000
05A1000a
05A2500
05A2500a
Part Number
05A2505
05A2505a
Power supply
Electrical connections
Measuring ranges
Be sure to observe chapter 5
External range indication
Remote range selection
Combined analog/digital indicator
Output signals
Response time for 90% FSD
Accuracy after calibration
Reproducibility
Influence of ambient temperature
Influence of barometric pressure
Influence of sample gas flow
Sample gas inlet pressure
Sample gas outlet pressure
Flow rate of sample gas
www.mc-techgroup.com
PMA 50 EX, power supply 230 V, pressure range 0.6 to 1.1 bar abs.
PMA 50 EX, power supply 115 V, pressure range 0.6 to 1.1 bar abs.
PMA 50 EX/P/PD-1-50, power 230 V, (not with SIL-certification)
pressure compensation 0.6 to 1.5 bar abs. with purging the enclosure via two
breathing and draining devices in the in- and outlet 1/4" NPT female
PMA 50 EX/P/PD-1-50, power 115 V, (not with SIL-certification)
pressure compensation 0.6 to 1.5 bar abs. with purging the enclosure via two
breathing and draining devices in the in- and outlet 1/4" female
PMA 50 EX/P/PD, power 230 V, (not with SIL-certification)
pressure compensation 0.6 to 1.1 bar abs.
PMA 50 EX/P/PD, power 115 V, (not with SIL-certification)
pressure compensation 0.6 to 1.1 bar abs.
230 V AC (standard) or 115 V AC (a) -15 % to +10 %, 40-60 Hz, 35.5 VA
Via Ex e connection box 3 x cable gland
cable diameter : 7 mm–13 mm (M20), 14 mm–18 mm (M25)
2
terminals 0.5 to 2.5 mm
, Tightening torque 0.6 Nm
(power, signals, range position and remote choice, status signal)
Selectable for 0-1, 0-3, 0-10, 0-30 and 0-100 vol% O
choice via turning selection switch at PMA 50 EX or remote switching
Potential free contact for all measuring ranges.
Switching capacity 48 V DC, 200 mA DC,
Minimum contact rating 5 V/1 mA
Measuring ranges selectable via potential free contacts max. 30 V DC, 3 mA DC.
The function is displayed at the PMA 50 EX via LED.
Analog meter with a scale of 0-30 and 0-100 % for each selected range
digital meter 4 ½ digit 9 mm high LCD-indicator for 0-100 % O
0.01 vol% O
2
0/4-20 mA burden 270 Ω for every measuring range, electrically isolated; output
voltage max. 15 V (ex works).
Switchable max. burden 800 Ω, output voltage max. 30 V.
Output current limiting adjustable 20 mA to 22 mA
20.5 mA ex works
0-10 V DC, burden >100 KΩ for range 0-100 % isolated.
< 5 seconds at 60 Nl/h air
 1 % of the span value of measuring range after calibration or  0.02 %O
ing on which value is the higher one
Analog output = < 1 % of measuring range / digital indication =
± 0.01 vol% O
2
No influence up to 50 °C [122 °F]
The oxygen reading varies in direct proportion to changes of the barometric
pressure.
No influence from 0.6-1.5 bar abs. at PMA 50 EX/P/PD(-1-50) with process pres-
sure compensation
Variation in gas flow between 0-60 Nl/h air will cause a difference of
< 0.2 vol.% O
2
Max. 1.1 bar abs.standard or in case of purging up to 1.5 bar abs.
(Minimum inlet pressure is required for necessary gas flow, PMA50 EX has no inte-
grated pump)
Outlet of analyzer must discharge freely into atmosphere
or 0.6 to 1.5 bar abs. at version PMA 50 EX/P/PD(-1-50) with process pressure com-
pensation
Min. 10 Nl/h up to max. 60 Nl/h adjustable externally via flow meter (no flowmeter
inside)
PMA 50 EX | 1.01.04
linear
2
reading, selectivity
2
depend-
2
11