Temperature of sample gas
O
-Transmitter temperature
2
Temperature cut-off
Ambient temperature
Storage temperature
Medium touched material
Connections measuring gas
Flow failure
Status signal outlet
Classification/Protection
Housing/color
Dimensions (H x W x D)
Weight
Certificate No.
10.1

OPTIONS

Part
Options
number
05A9015
Zero suppression in range of 1 to 97 vol%; (Only in connection with Part No. 05A2500(a))
Externally switchable via potential free contacts, 30 V DC, 3 mA DC
(not with SIL-certification)
05A9005
Extra charge for one breathing and draining device for PMA50 EX enclosure, for sample gas pressures up
to max. 1.5 bar abs. and non-corrosive gases
05A9000
Purging the enclosure via two breathing and draining devices in the gas inlet and outlet, for sample gas
pressures up to max. 1.5 bar abs and/or corrosive gases, connection 1/4" NPT female, purge gas inlet pres-
sure max. 1.1 bar abs., flow rate 10-60 Nl/h
90A0009
Measuring cell type PMC-1LB, solvent resistant.
90A0006
Measuring cell PMC-1G with glass solder. O-ring made of Chemraz
Please note: Nl/h and Nl/min refer to the German standard DIN 1343 and are based on these standard conditions: 0 °C [32 °F],
1013 mbar.
®
Chemraz
is a registered trademark for perfluoroelastomer by Greene Tweed, USA.
12
-10 up to +50 °C [14 up to 122 °F] dry gas
Fixed at +55 °C [131 °F] (factory-set)
At 72 °C [161.6 °F] via thermal fuse, non-reversible
0 to +50 °C [32 to 122 °F]
-20 to +60 °C [-4 to 140 °F] at relative air moisture of 0-90 % r.h.
Platinum, glass, PTFE, PVDF, stainless steel 316Ti, Epoxy
1/4" NPT female
Thermo-conductive flow sensor downstream mounted after measuring cell
Change-over contact, switching capacity 250 V AC, 2A AC, 48 V DC, 200 mA
DC minimum contact rating 50 mW for temperature <+45 °C [113 °F]/> +60
°C [140 °F], defect light beam, measuring cell not coupled, flow failure
< 5 Nl/h, power supply error control, mains voltage breakdown, failure meas-
uring range selection
II 2G Ex db eb IIC T4 Gb
IP54 EN 60529
EX-d e wall mounted explosion-proof housing/white
475 (535 mm with breathing and draining device) x 355 x 200 mm [≈ 18.7"
(535 mm [≈ 21.1"] with breathing and draining device) x 14" x 7.9"]
Approx. 22 kg [≈ 48.5 lbs]
IBExU 16 ATEX 1192
IECEx IBE 16.0041
PMA 50 EX | 1.01.04
®