Battery; Environmental Considerations (Ambient); Laser Safety Designation (Iec 60825-1); Table 10: Table Size Requirements - Daytona P200T User Manual

Optos panoramic ophthalmoscope
Table of Contents

Advertisement

Chapter 4 - Technical Specification
4.1.6 Table requirements
Do 
plug the scan head into a power block or powered table. The scan head must be
NOT
plugged directly into a wall socket.
The scan head should be positioned in the middle of the table width. The front of the scan
WARNING
head should be inline with the front of the table to allow the patient to get into the correct
alignment position.
The scan head and touch screen must be positioned on a table that meets the necessary safety standards.
The table
support the weight of the scan head.
MUST

TABLE 10: Table size requirements

Dimension
Details
Width
887 mm (35 inches) (not including wheel position)
Depth
623 mm (24.5 inches) (not including wheel position)
Height
725 mm to 1205 mm (29 inches to 47 inches)
Where wheelchair accessibility is required the table must meet wheelchair accessibility standards.

4.1.7 Battery

Do
remove the scan head covers. Only Optos trained technical personnel are
NOT
permitted to service the device.
CAUTION
The device contains a 3V Lithium coin battery.

4.1.8 Environmental considerations (ambient)

Do
operate the device in an environment that exceeds the environmental limits, see Warnings and
NOT
cautions on page 11.

TABLE 11: Environmental ranges

Status
Normal use
Storage
Transport
Note
Device is DEHP free.

4.2 Laser safety designation (IEC 60825-1)

All P200T product variants are Class 1 laser devices at the eye, and comply with IEC/EN
60825-1:2007 and 21 CFR1040.10 and 1040.11 except for deviation pursuant to Laser
Notice 50, dated June 24, 2007.
CAUTION
Based on current scientific knowledge, a Class 1 laser device can be considered as safe
by engineering design, and safe under reasonably foreseeable conditions of operation.
Page 50 of 60
English
Temperature
+10°C to +35°C
-10°C to +60°C
-10°C to +60°C
Relative Humidity
(Non-condensing)
30% to 75%
10% to 95%
10% to 95%
Copyright 2021, Optos plc. All rights reserved.
Atmospheric Pressure
700hPa to 1060hPa
700hPa to 1060hPa
500hPa to 1060hPa
Part Number: G102748/11_ENG