Chapter 3 - How to...
c. Turn their head slightly so that their nose is on the outside of the face pad. You should check
that the patient's head is vertical and has not been tilted.
5. After they are comfortable, ask the patient to move to the correct alignment position. The alignment
colors seen by the patient are also shown on the touch screen.
6. Adjust the alignment. The alignment target will change to green when the scan head is the correct
distance from the patient's eye. If the target turns red the scan head is too close. If the target turns
blue the scan head is too far away.
When aligning the patient the center of the cross hairs should be close to the center of the pupil
l
as possible.
In some instances you may choose to capture an image when the patient is aligned to the blue
l
or red target. If the patient has an
target is red. If the patient has very large pupils you may choose to capture when the target is
blue.
7. The x, y, and z alignment co-ordinates are shown. Ideally these should be as near to zero as practical,
depending on the patient being imaged.
8. Capture the required image modality from the following list:
Reference

How to capture SLO images

How to capture eye steered images
How to capture stereo images
3.2.4 How to capture SLO images
To capture SLO images use the procedure below. You have the option to capture the following image
modalities:
optomap
l
optomap
plus
l
optomap
af
l
1. Prepare the device for image capture, see How to prepare for image capture on the previous page.
2. Align the patient, see How to align the patient on the previous page.
3. Ask the patient to open both eyes wide, and to look at the center of the alignment target. Using the live
SLO image, center the cross hairs on the pupil until the target is aligned and turns green.
4. Capture the image either manually or using auto-capture.
Auto-capture: Auto-capture is enabled when
l
The system captures an image when the scan head is correctly aligned. You can press the
screen to manually capture the image.
Manual capture: Align the scan head into the correct position. Press the screen to capture the
l
image when the pupil is in the center of the cross hairs and the patient alignment target is
green.
1. Intraocular lens
Page 34 of 60
English
IOL
1
or small pupils you may choose to capture when the
Automatic
is shown in green on the touch screen.
Part Number: G102748/11_ENG
Copyright 2021, Optos plc. All rights reserved.
34