UMS T8 User Manual page 32

Long-term monitoring tensiometer
Table of Contents

Advertisement

T8 configuration with tensioVIEW®
bus address plus the sub address. For the default value depth = 1
the internal sensor readings are used.
Associate depth DACB
Associate a depth to reading DACB.
DAC error output
Error output value for DAC [mV]. If an incorrect reading occurs (e. g.
filling status insufficient) this voltage value will be put on the analog
output.
DAC output range
Set the output range of the analog reading to 0 to 1 V; 0 to 2 V or 0
to 5 V. A sensor reading will be amplified within this range, for
example -100 kPa to +100 kPa can correspond to 0 to 2 V.
DAC temperature output range
Output range for temperature.
DAC pressure output range
Output range for pressure.
Digital output
Digital output power saving
Activate the power saving function of the digital output. Any burden
connected to the digital output will be powered shortly every 5
seconds (e. g. a LED will blink every 5 s indicating a malfunction)
Function of the digital output
Evaluation through a table. If the digital output is used for control of
the filling status, 1 indicates a poor filling status.
If used as a comparator the value of a reading is compared to the set
threshold (see "Lower threshold"). If used as a window comparator
the reading is compared with both threshold values. In case the
reading is in between both threshold values the digital output is set to
1.
Associate reading to digital output
This is the relevant reading for the comparator.
Associate depth to digital output
The depth value for comparator, 1 = individual sensor (see
"Associate  depth  DACA")
Invert digital output function
If enabled all digital output functions will be inverted.
Lower threshold integer
Integer part of the lower threshold or window threshold value.
32/58