Tensiometer T8
5. Push down the shaft water retaining disk to a position directly on
the soil surface.
6. Slide on the thermal insulation tube over the capillary filling tubes.
7. If the plug is not connected right away leave the protective cap on
the plug. Water tightness is only assured when the plug is kept
clean.
8. Connect the Tensiometer signal wires as specified (see chapter
„Connecting  the  T8")
The T8 can be connected to:
data loggers with analog input or serial RS485 or SDI12,
a PC or laptop via tensioLINK bus as a standalone sensor or
within a sensor network,
an additional 6 V Battery to run the T8 in the auto-logging
function (internal data logger, requires tensioVIEW software and
USB converter for configuration),
the INFIELD7 for taking spot readings or downloading stored
readings.
Please note:
Especially in loamy, clayey soils a high pressure can occur just by
inserting the T8 into the borehole. Thus, the pressure values
should be continuously observed during installation, either with the
INFIELD7 handheld or online with laptop, USB Converter and
tensioVIEW.
Note for the latter that the readings are refreshed only every 5
seconds by the internal microprocessor.
The less air is inside the cup and the better the soil's conductivity
is, the faster the Tensiometer will respond to tension changes.
If the soil is dryer than -90 kPa, it does not make sense to refill the
Tensiometer. The refilling procedure will be done best at the time
when the Tensiometer installed in the next lower level has reached
the value of drying off of the upper Tensiometer.