General Information The NuFlo TMP-100 Turbine Meter Pulse Divider Circuit is a solid-state electronic instrument designed for use with a NuFlo turbine flowmeter. The TMP-100 provides a pulse output in the form of two dry contacts from a two-pole relay, single-pole relay or an opto- isolated open collector transistor output.
Installation Mounting The TMP-100 circuit board is designed to be mounted with screw-mounted standoffs or push-on style circuit board supports. Mount the TMP-100 in a clean, dry location suitable for electronic circuit assemblies where it will not be subjected to extreme operating temperatures (below -20ºF or above +140ºF). The mounting location should ensure easy access to the terminal strip, divisor switches, and pulse duration adjustment.
Wiring All field wiring is attached to the instrument on the 16-position terminal strip near the edge of the circuit board (Figure 2). The power requirement for the unit is 10 to 28 VDC. 10 to 28 VDC power Turbine meter magnetic pickup Turbine meter frequency out...
Installations requiring a turbine meter frequency output are labeled A&S and COM. These terminals provide a 0- to 5-V square-wave output, representing the turbine meter frequency. A load with a resistance 250K ohm or larger may be connected to these terminals. The shield wire must be connected to the COM terminal.
Page 10
Determining the Divisor The divisor is determined by the flowmeter calibration factor and the conversion factor for the desired units of calibration. Example: A customer has a 2-in. turbine meter with a calibration factor of 55.2 pulses per gallon. The desired unit of registration is barrels.
Switch SW1 positions 1 through 10 and switch SW2 positions 1 through 9 are used to select a divisor. Switch SW2, position 10, is used for pulse direction. This function is described in the Pulse Direction and Duration section. Determining Switch Position To determine which switches should be closed, perform the following steps: 1.
Adjusting the pulse duration requires an oscilloscope, a power supply, and a signal generator. To adjust the pulse duration, perform the following steps: 1. Connect the power supply to the POWER IN terminals as shown in Figure 2. 2. Connect the signal generator in place of the turbine meter on the TM IN terminals. 3.
Ordering Information Circuit Assemblies Part No. Description 9A-100005170 TMP-100 with two-pole relay and circuit board supports 9A-100079944 TMP-100 with single-pole relay and circuit board supports 9A-100079897 TMP-100 with opto-isolated module and circuit board supports Circuit Board Option Part No. Description Rate option - 4-20 mA analog output representing flow rate 9A-100005167 (specify flowmeter size or full scale frequency when ordering)
Page 15
WARRANTY - LIMITATION OF LIABILITY: Seller warrants only title to the products, software, supplies and materials and that, except as to software, the same are free from defects in workmanship and materials for a period of one (1) year from the date of delivery. Seller does not warranty that software is free from error or that software will run in an uninterrupted fashion.