Firmware Upgrades And Flash Signature Errors; Upgrading The Firmware; Firmware Signature Errors; Checking The Firmware Version And Signature - Campbell SDM-SIO1A Instruction Manual

Serial input/output module
Hide thumbs Also See for SDM-SIO1A:
Table of Contents

Advertisement

SDM-SIO1A Serial Input/Output Module
5

Firmware upgrades and flash signature errors

5.1 Upgrading the firmware

5.2 Firmware signature errors

5.3

Checking the firmware version and signature

22
In the unlikely event that the operating system of the SDM-SIO1A needs to be
upgraded. This can be achieved by connecting it to a PC and using the Campbell
Scientific Device Configurator program to load the operating system. (This
program can be downloaded free of charge from
www.campbellsci.com/downloads).
The SDM-SIO1A needs a reliable source of 12V power connected in the standard
way and a cable made similar to that shown in Section 3.3.7 to connect to an RS-
232 port on the PC, except there is no requirement to connect any handshaking
lines.
Start the Device configurator program and select the SDM-SIO1A device (the
SDM-SIO1A device may not be available in older versions of the software. New
versions can be downloaded from the Campbell website). Follow the instructions
for that device to load a new operating system. At the end of the process a
success message will be shown if successful. During the loading of the operating
system do not disturb or disconnect power to the SDM-SIO1A otherwise it may
need to be returned to the factory for repair.
The operating system is stored in "flash" memory. When a new version is loaded
a checksum signature is automatically created the first time its run and stored in
memory too. This signature value can be read back using the logger and compared
to the signature supplied with the version that was just loaded (contact Campbell
Scientific if you require to do this).
If the SDM-SIO1A is not operational or is exhibiting random faults then the flash
memory may be corrupted. This is a very unlikely event due to the robust nature
of the flash device used. The unit automatically checks the flash memory against
its signature upon power up. If an error is found it will send out the string 'sig
error:XXXX:YYYY' when an error is detected, where 'XXXX' is the signature as
it's being read and 'YYYY' is the stored signature.
The error string is sent out via the RS-232 port automatically and, will also be
returned to the datalogger the next time any attempt is made to read information
from the sensor. The message will be output at the default baud rate of 9600bps,
8-bits, 1 stop and no parity.
If this error is seen contact Campbell Scientific to obtain a copy of the latest
operating system and load it into the SDM-SIO1A using the above procedure. If
this does not correct the fault then the unit may be faulty and will need to be
returned to the factory for repair.
It is not possible to check the current firmware version using the device
configuration program as the SDM-SIO1A is designed as a transparent RS232
device. Instead the firmware version can be checked by loading a program into the
logger to which it is attached. The program below shows use of the SDMGeneric
command to ask the SDM-SIO1A with a specific address what the firmware
version is (Ver_value) and the signature of the firmware (Sig_value).