5. Connect the tube to the inlet port and use the tube fixture to secure
the tube against the foot.
6. Make sure that the pump is functioning properly and the system is
not leaking.
FINISHING / LINER OPTIONS
Össur recommends that the Unity™ System is used with the Iceross Seal-
In V liner.
caution: Make sure the Flex-Foot sock does not interfere with the Unity™
System as it could compromise the function of the pump.

MAINTENANCE

The following components of the Unity™ System might need cleaning or
replacing during the lifetime of the product. This is to be considered
normal wear and is not covered by the Limited Product Warranty.
• Membrane
• Duckbill valves
• Tube
It is recommended to have the product inspected every 6 months by a
clinical specialist.
Clean the components with mild soap and rinse with water, isopropyl
alcohol or Iceross Clean&Simple spray.
DO NOT use solvents stronger than isopropyl alcohol.
Please refer to the current Össur printed catalog or online pages via
www.ossur.com for replacement parts available and appropriate item
numbers.
IN CASE THE UNITY™ SYSTEM IS NOT FUNCTIONING PROPERLY
Check the following:
• Is there a leakage in the socket?
• Is the tube punctured?
• Is the membrane punctured or torn?
• Has the membrane been pulled out of the membrane housing?
• Is there dust or grit inside the system?
• Is the tube fully inserted on the inlet port? (Check both at the Unity™
Socket valve and at the Unity™ Membrane housing)
• Are the inlet and outlet ports tightened?
• Is the Unity™ Socket valve leaking?
• Is the sock inside the foot cover affecting the systems mechanical
movements?
AFTER DISASSEMBLY / ASSEMBLY
Check the following:
• Are the duckbill valves oriented in the correct direction (Figure 7)?
• Is the membrane fully inserted into the housing?
• Are the inlet and outlet ports tightened?
• Is the tube fully inserted on the inlet port? (Check both at the Unity™
Socket valve and at the Unity™ Membrane housing)
THE TUBE
The tube is the most sensitive part of the Unity™ System and special
attention should be paid to it in case of any problems.
Make sure the tube is NOT:
• Folded (Figure 8).
• Pressed up against sharp objects.
• Punctured.
• Dangling during gait.
• Located laterally, as there is more likelihood of damage.