Disconnect the battery charger and move the Product or battery to a cool, non-
flammable location if the rechargeable battery becomes hot (>50 °C) during the
charge period.
Replace the rechargeable battery after 5 years of moderate use or 2 years of
heavy use. Moderate use is defined as recharged twice a week. Heavy use is
defined as discharged to cutoff and recharged daily.
To replace the battery, send the product to an authorized Fluke Networks Service
Center.
To get the best performance from the lithium-ion battery:
Do not charge the Product for more than 24 hours as a reduced battery life may result.
Charge the Product for at least 1.5 hours every 6 months for maximum battery life. Without
use, the battery will self-discharge in approximately 6 months.
Figure 9
shows how to charge the battery.

Product Specifications

For complete Product Specifications go to our website.
Figure 9. Charge the Battery