Product Specifications - Braun ThermoScan IRT6510 Manual

Hide thumbs Also See for ThermoScan IRT6510:
Table of Contents

Advertisement

Available languages

Available languages

Product specifications

Displayed temperature range:
Operating ambient temperature range:
Storage temperature range:
Operating and storage relative humidity:
Display resolution:
Accuracy for displayed temperature range
• 35 °C – 42 °C:
• Outside this range:
• Clinical repeatability:
Battery life:
Service life:
This thermometer is specified to operate at 1 atmospheric pressure or at altitudes with an
atmospheric pressure up to 1 atmospheric pressure (700-1060hPa).
See Instruction for use
Equipment with type
BF applied parts
Subject to change without notice.
This appliance conforms to the following standards:
Standard Reference Edition Title:
EN 12470-5: 2003 Clinical thermometers - Part 5: Performance of infra-red ear thermometers (with
maximum device).
EN 60601-1: 2006 Medical electrical equipment – Part 1: General requirements for basic safety and
essential performance.
EN ISO 14971: 2012 Medical devices – Application of risk management to medical devices.
EN ISO 10993-1: 2009 Biological evaluation of medical devices – Part 1:
Evaluation and Testing within a risk management process.
EN 60601-1-2: 2007 Medical electrical equipment – part 1-2: General requirements for basic safety
and essential performance
– Collateral standard: electromagnetic compatibility
– Requirements and tests
EN 980: 2008 Symbols for use in labeling of medical devices.
EN 1041: 2008 Information supplied by the manufacturer of medical devices.
EN 60601-1-11: 2010 Medical electrical equipment -- Part 1-11: General requirements for basic
safety and essential performance -- Collateral standard: Requirements for medical electrical
equipment and medical electrical systems used in the home healthcare environment.
This product conforms to the provisions of the EC directive 93/42/EEC.
Medical electrical equipment needs special precautions regarding EMC. For detailed description of
EMC requirements please contact your authorized local service centre (See warranty card for details).
Portable and mobile RF communications equipment can affect Medical electrical equipment.
Please do not dispose of the product in the household waste at the end of it useful life.
To protect the environment, dispose of empty batteries at appropriate collection sites
according to national or local regulations.
34 – 42.2 °C
10 – 40 °C
-25 – 55 °C
10 – 95 %RH (non condensing)
0.1 °C
Maximum Laboratory Error
± 0.2 °C
± 0.3 °C
± 0.14 °C
2 years / 600 measurements
5 years
40 °C
10 °C
-25 °C
Operating temperature
Storage temperature
55 °C
Keep it dry