Errors and troubleshooting

Error message
10
Situation
No Braun lens filter is attached.
The thermometer probe was not
positioned securely in the ear.
An accurate measurement
was not possible.
POS = position error
Ambient temperature is not within
the allowed operating range
(10 – 40 °C).
Temperature taken is not within
typical human temperature range
(34 – 42.2 °C).
HI = too high
LO = too low
System error – self-check display flashes
continuously and will not be followed by
the ready beep and the ready symbol.
If error persists,
If error still persists,
Battery is low, but thermometer will still
operate correctly.
Battery is too low to take correct
temperature measurement.
Solution
Attach new, clean Braun lens filter.
Take care that the positioning of the
probe is correct and remains stable.
Replace Braun lens filter and reposition.
Press Start button to begin a new
measurement.
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between 10 and
40 °C .
Make sure the probe tip (pre warm tip)
and lens are clean and a new, clean
Braun lens filter is attached. Make sure
the thermometer is properly inserted.
Then, take a new temperature.
Wait 1 minute until the thermometer
turns off automatically, then turn
on again.
... reset the thermometer by re-moving
the batteries and putting them back in.
... please contact the service centre.
Insert new batteries.
Insert new batteries.