Pad Configuration; Unlatch Coil; Positioning The Patient - GE 5160986-2 Operator's Manual

1.5t and 3.0t 8ch wrist array coils
Table of Contents

Advertisement

March 2022

5.2.3 Pad configuration

Six pads are supplied with the Wrist Array.
The base pad (1) provides patient comfort.
The palm (2), upper (3), wrist (4), and finger (5) pads support the
hand and wrist to help prevent motion artifact.
The elbow pad (6) supports and stabilizes the forearm for patient
comfort and to help prevent motion artifact.
*Pads may be grey.

5.2.4 Unlatch coil

Unlatch the coil by pushing the lock toward the cable end of the coil
and pulling the coil open, toward you.

5.2.5 Positioning the patient

Position the patient with feet first, with their hand and wrist fitting
the contour of the coil.
Place the elbow pad under the patient's arm to help keep the wrist
straight and use the other pads to ensure the patients hand is
immobilized and to increase patient comfort.