Granular Activated Carbon and HEPA Filter Replacement:
1. Use an appropriate stepladder release the two bungee cords securing the
granular activated and carbon and HEPA filters, as shown in Figure 6-3.
2. When handling used filters or prefilters, always wear appropriate personnel
protective equipment.
3. Lift the granular activated carbon and HEPA filters straight up and out of
the cabinet top.
4. Install the replacement filters, with the HEPA filter below the granular
activated carbon filter into the cabinet. The gaskets on both filters should be
on the lower (bottom) surface.
5. Reinstall both bungee cords to secure the filters in place. The unit is now
ready to resume operation.
6. Filters should be disposed of properly. HEPA Filters should be treated as a
biohazard.