The common service operations necessary to maintain the Evidence Drying
cabinet for peak performance are listed below.

Routine Maintenance Schedule

Weekly
 Using 70% ethanol, or a suitable disinfectant, surface disinfect the inside of
the cabinet.
 Using an appropriate glass cleaner, or LabSolutions Glass & Surface
Wipes, Labconco part # 1570000; clean the door and the surface of the UV
lamp.
 Operate the cabinet blower, noting if the Filter System Light turns on.
 Ensure the cabinet water shut off valve is in the off position.
Monthly (or more often as required)
 Using a damp cloth, or LabSolutions Glass & Surface Wipes, Labconco
part # 1570000; clean the exterior surfaces of the cabinet, particularly the
front and top of the cabinet, to remove any accumulated dust.
 All weekly activities.
 Check the prefilters; replace both if needed.
Annually
 All monthly activities.
 Replace HEPA and granular activated carbon filters.
26
:
:
n
g
t
h
e
C
a
n
g
t
h
e
C
a
b
i
n
e
t
b
i
n
e
t
Product Service 1-800-522-7658