Symptom
No Detection
Moisture Collected In
Detector or Flow System
Tubing
Page 76 of 92
Probable Cause
No high-voltage
Soot in System
Gas leak
No detector electronics
response
Defective Supply
Membrane Dryer leaking
internally
Water in Reaction Chamber
Membrane dryer(s) is bad
Dryer flow is too low
Corrective Action
Replace high-voltage supply.
Check for soot deposits in
system.
Use gas leak test to find leaks
and repair as needed.
Cause a light leak in the
detector to determine
operation of electronics.
Check all supply outputs and
replace defective supply.
Replace Membrane Dryer.
Remove all water from
chamber.
Replace membrane dryer.
Check dryer flow meter and
increase flow as necessary.