PAC R6000S Operation Manual page 75

Table of Contents

Advertisement

Symptom
Response Changing
Nonlinear Response
PAC L.P., All Rights Reserved, 2015
Probable Cause
Gas leak
Contaminated gas supply
Contaminated Pyrotube
Unstable AC power source
Unstable Gas Regulator
Reduced UV source output
PMT Voltage Change
Abnormal ambient temp
change
Light leakage
Defective PMT or Cooler
Defective Pre-amp PCB or
Amplifier PCB
Amplifier saturation
Contaminated gas supply
Contaminated Pyrotube
Incomplete sample
combustion
Reduced UV source output
PMT Voltage Change
R6000S Operation Manual: Rev. G
Corrective Action
Use gas leak test to find leaks
and repair as needed.
Replace gas supply.
Clean or replace Pyrotube.
Determine cause of instability
and correct.
Replace Regulator.
Replace UV Lamp.
Replace or adjust PMT.
Stabilize ambient temperature.
Use light leak test to find leaks
and repair as needed.
Replace PMT or Cooler.
Replace Pre-amp PCB or
Amplifier PCB.
Reduce sample concentration,
or amplifier gain.
Replace gas supply.
Clean or replace Pyrotube.
Ensure furnace temperature
and gas flows are correct.
Replace UV Lamp.
Replace PMT.
Page 73 of 92