Calibrating The Analyzer; Set Zero - PAC R6000S Operation Manual

Table of Contents

Advertisement

R6000S Operation Manual: Rev. G
g. Set backpressure regulator to 5 - 9 psi. (This pressure setpoint is not critical. The flow setting in
the next step is).
h. Adjust flow meter to 1.0.
i.
Readjust backpressure regulator back to the pressure set in step 7E, if necessary.
j.
Set Sample/Cal switch to Sample and Zero/Span switch to Zero.
k. Set Zero/Span switch to Zero.
l.
Monitor flow meter for 1.0. If flow meter does not read 1.0, adjust regulator at cylinder.
m. Set Zero/Span switch to Span.
n. Monitor flow meter for 1.0. If flow meter does not read 1.0, adjust regulator at cylinder.
o. Set Sample /Cal switch to Sample and Zero/Span switch to Zero.
p. Adjust flow meter to 1.0.
q. Set backpressure regulator to the pressure set in step 7g.
NOTE: The sample/calibration system is now adjusted. When switching between sample and calibration
gases the flow meter and backpressure should maintain the same value.
3.2

Calibrating the Analyzer

NOTE: Ensure all gas supplies are turned ON and are at the correct flow and pressure settings and that
an adequate calibrating standard is available.
NOTE: Make sure the instrument has been running for at least thirty (30) minutes after all settings have
been achieved before making any adjustments.
3.2.1

Set Zero

1. Set sample/calibrate switch to calibrate; set zero/span switch to zero.
2. Verify Sample flow meter at 1.0.
NOTE: The sample flow meter will allow sample, span and zero gases to flow through when the
appropriate solenoids are switched/activated.
3. After detector output has stabilized for five minutes (shown by a consistent reading on the level meter
or an auxiliary recorder), press the auto zero button.
4. Once the auto zero light goes out, observe that the detector maintains 0.0 on the level meter or 4.0
mA on the current loop for two minutes.
Page 44 of 92
Do not attempt to operate the analyzer until all installation procedures outlined
in the previous sections of this manual have been successfully completed.
Prior to normal operation, following heading, become familiar with the location
and use of all controls, indicators, connections, and accessories. Also,
carefully read all instructions and warning labels prior to operating any portion
of the system.
PAC L.P., All Rights Reserved, 2015

Advertisement

Table of Contents
loading

Table of Contents