R6000S Operation Manual: Rev. G
Figure 4.
Signal Out Terminal Block: Detector Module
2. Connect a power cord from the detector module to an adequate supply (greater than 5 Amps@ 115-
VAC or 3A @ 230-VAC).
3. Connect a power cord from the furnace module to an adequate supply (greater than 10 Amps @ 115
VAC or 5 A @ 230 VAC).
If a single circuit is used to supply both modules, it must be rated greater than 20 Amps @115 VAC or 10 Amps @ 230
VAC.
Page 38 of 92
POSSIBLE ELECTRIC HAZARD!
DEATH OR SERIOUS INJURY COULD OCCUR IF NOT AVOIDED.
Electrical shock and equipment damage or failure may result due to improper
safety grounding. Without a proper equipment ground, for example a broken or
faulty grounding conductor, all normally grounded conductors and surfaces
may become energized and potentially hazardous.
HOW TO PROCEED: Avoid possible hazard by ensuring all grounds are
properly grounded. Ensure that the green safety ground wire from the power
input cable and all other ground wires are securely fastened to the ground
terminal screw on the chassis, following figure.
PAC L.P., All Rights Reserved, 2015