Parameterization Via The Local Operating Interface; Operating Instructions - ABB SensyMaster FMT230 Commissioning Instruction

Hide thumbs Also See for SensyMaster FMT230:
Table of Contents

Advertisement

SENSYMASTER FMT230, FMT250 THERMAL MASS FLOWMETER | CI/FMT230/250-EN REV. A
5.4.2

Parameterization via the local operating interface

DANGER
Risk of explosion during operation of the device with open
terminal box!
Only perform parameterization of the device via the local
operating interface outside the potentially explosive area!
A PC / notebook and the USB interface cable
(3KXS310000L0001) are required to configure the device via the
device's local operating interface.
In conjunction with the HART-DTM and the software "ABB
AssetVision" available at www.abb.com/flow, all parameters
can also be set without a Modbus connection.
1
Fig. 26:
Connection to the local operating interface
1 Local operating interface 2 Programming plug
3 PC / notebook 4 USB interface cable
1.
Open device terminal box.
2.
Connect programming plug to the local operating interface
of the device.
3.
Insert USB interface cable into a free USB female connector
on the PC / notebook.
4.
Switch on the device power supply.
5.
Start ABB AssetVision and perform the parameterization of
the equipment.
Detailed information on operating the software is available in
the relevant operating instructions and the DTM online help.
2
3
4
G11625
5.5

Operating instructions

When operating the device, please note the following:
— Aggressive or corrosive media may lead to the damage of
wetted parts of the sensor. As a result, measuring medium
under pressure can leak out.
Measuring medium under pressure can leak out due to
fatigue on the gasket of the sensor connection or the
process connection (e.g. flange or pipe fitting).
If there is a chance that safe operation is no longer possible,
take the device out of operation and secure it against
unintended startup.
25