Verifying Acceptable Spatial Resolution, Artifact Levels and
Low Contrast Visibility
To verify acceptable spatial resolution, artifact levels and low contrast
visibility, proceed as follows:
1. Print the QC Test image or use the previously printed QC Test image used
to establish the daily operating density levels.
2. Visually check the QC test image for artifacts: no significant disturbing
artifacts should be visible.
3. Check the spatial resolution in each of the three circles. Within each circle
there are three groups, each consisting of five dots. All five dots of each
group must be visible with a magnifying glass. The smallest cluster of 5
dots is only visible if the viewing conditions are good.
4. Check the Low Contrast Visibility at both the high (100 / 95%) and low
end (0 / 5%) of the density scale. You should be able to see the circle in
the square and the upper circle.
5. Record these values at the top of Chart 3 ('Artifacts and Spatial Resolution
Control Chart').
6. These charts will be used for the weekly quality test.
Related Links
QC test image
Charts for general radiography QC
Performing the weekly QC tests

Preventive maintenance schedule

Drystar AXYS XL | Advanced Operation (key-operator mode) | 101
WARNING:
Good viewing conditions are important for the correct
interpretation of both diagnostic and test images. Make sure
that the light box intensity (luminance) is between 2000 and
4000 cd/m² (4500 and 6500 °K). Use a magnifying glass and use
shutters to collimate. Make sure the ambient light is low.
WARNING:
In case of significant artifacts or insufficient spatial resolution,
the reason must be found and the problem solved before any
further clinical films can be printed.
on page 95
on page 103
on page 120
on page 139