3. Record these values as measured dimensions A and B on Chart 4
('Geometric Consistency Control Chart').
4. Compare the measured A and B dimensions with the reference dimension
values, A
The differences between measured dimensions of A and B and the
reference values A
5. Check for image distortion.
6. Calculate the aspect ratio by dividing A by B.
The result must be 1 +/- 0.01
Related Links
Establishing the image geometry reference values
Drystar AXYS XL | Advanced Operation (key-operator mode) | 105
and B
on Chart 4 ('Geometric Consistency Control Chart').
ref
ref
and B
ref
ref
WARNING:
If the image size or distortion values are outside of limits,
contact your local service organization to resolve the problem.
should be less than or equal to 1.0%.