Remarks For Hf-Emission And Immunity - AGFA Drystar AXYS XL User Manual

Hide thumbs Also See for Drystar AXYS XL:
Table of Contents

Advertisement

Remarks for HF-emission and immunity

This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
RF Emission
Measurements
High frequency
RF emissions in
accordance with
CISPR 11
High frequency
RF emissions in
accordance with
CISPR 11
Harmonic emis-
sion in accord-
ance with IEC
61000-3-2
Voltage fluctua-
tions / flickering
in accordance
with IEC
61000-3-3
The device is used in a professional healthcare / radiological environment.
Environmental conditions are stated in the user manual.
This device was tested for a professional healthcare environment as described
above. Nevertheless the HF-emission and immunity can be influenced by
connected data cables depending on length and the manner of installation.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Resistance to
Jamming Test
Discharge of
static electricity
Drystar AXYS XL | Remarks for HF-emission and immunity | 129
Agree-
Electromagnetic Environment Guidelines
ment
Group 1 The device uses high frequency energy exclu-
sively for its internal functions. For this reason,
its high frequency RF emission is very low and
it is improbable that neighboring electronic
equipment will be disrupted.
Class A
The emissions characteristics of this equipment
make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a
residential environment (for which CISPR 11
class B is normally required) this equipment
Class A
might not offer adequate protection to radio-
frequency communication services. The user
might need to take mitigation measures, such
as relocating or re-orienting the equipment.
Fulfilled
IEC 60601-1-2:2014
Test Level
+ 8 kV contact dis-
charge
Level of
Electromagnet-
ic Environment
Agreement
Guidelines
+ 8 kV contact
Floors should
discharge
consist of wood,
3752B EN 20210601 1655